Energy window setting for optimum Tl-201 cardiac imaging [Persian]

Document Type : Original Article

Authors

1 Department of Medical Physics, Tarbiat Modarres University, Tehran, Iran

2 Nuclear Medicine Department, Shahid Rajaie Heart Hospital, Iran University of Medical Sciences, Tehran, Iran

3 Department of Medical Physics, Lorestan University of Medical Sciences, Khoramabad, Iran

Abstract

  Introduction: Poor sensitivity and poor signal to noise ratio because of low injected thallium dose and presence of scattered photons are the main problems in using thallium in scintigraphic imaging of the heart. Scattered photons are the main cause of degrading the contrast and resolution in SPECT imaging that result in error in quantification. Thallium decay is very complicated and photons are emitted in a wide range of energies of 68-82 keV. It seems possible to achieve better primary to scattered radiation ratio and better image sensitivity simultaneously if the energy window setting is carefully selected. Methods: This investigation was performed in three steps: Monte Carlo simulation, phantom experiment and clinical study. In simulation step, the new 4D digital NCAT phantom was used to simulate the distribution of activity (201Tl) in patient torso organs. The same phantom was used to simulate the attenuation coefficient of different organs of the typical patient's body. Two small defects on different parts of left ventricle also were generated for further quantitative and qualitative analysis. The simulations were performed using the SimSET simulator to generate images of such patient. The emissions arising from Tl-201 decay were simulated in four steps using the energies and relative abundances. Energy spectra for primary and scatter photons were calculated. Changing the center and width of energy windows, optimum energy window characteristics were determined. In next step jaszczak phantom was prepared and used for SPECT imaging in different energy windows. In last step SPECT images of 7 patients who had angiographic data were acquired in different energy windows. All of these images were compared qualitatively by four nuclear medicine physicians independently. Results: The optimum energy window was determined as a wider asymmetric window (77keV30%) that its center is not placed on photo-peak of energy spectrum. This window increased the primary counts rate and PTSR considerably as compared with the conventional symmetric energy window (67keV%). In a comparison which performed between clinical images acquired in suggested 77-30% window with conventional 67-20% window, a considerable increase was found in myocardial to defect contrast (1.541±0.368) and myocardial to cavity contrast (1.171± 0.099). A negligible increase was also found in total counts of images using this window. Conclusion: We found that conventional symmetric energy window (67keV± 10%) couldn't be a suitable choice for thallium heart imaging; furthermore three energy windows, 73keV-30%, 75keV-30% and 77keV-30%, were determined as optimum window options. For further analysis the images from such windows were compared in each three steps of this investigation. In all steps conventional symmetric energy window (67keV-20%) was introduced as the worst case and the asymmetric 77keV-30% was determined as the most suitable.

Keywords

Main Subjects