Comparative analysis of quality control results of 22 dual-energy X-ray absorptiometry (DEXA) systems across six manufacturers using a comprehensive one-day protocol

Document Type : Original Article

Authors
1 Department of Nuclear Engineering, Shiraz University, Shiraz, Iran
2 Department of Nuclear Medicine, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
10.22034/irjnm.2026.130431.1734
Abstract
Introduction: International societies involved in bone mineral densitometry emphasize the necessity of regular quality control (QC) for dual-energy X-ray absorptiometry (DEXA) systems. However, the current recommended procedures are not comprehensive and do not assess certain critical performance aspects—such as system stability under varying soft-tissue thickness and differing attenuation coefficients—that may affect measurement reliability. This study introduces and applies a one-day QC protocol designed to evaluate five key aspects of DEXA performance: in-vitro short-term precision (IV-STP), repeatability, stability across varying soft-tissue thickness (St-Thi), stability across differing attenuation coefficients (St-Att), and accuracy.
Methods: Twenty-two DEXA scanners from six manufacturers were evaluated using a lumbar phantom with Plexiglass slabs or a water tank to simulate soft-tissue variations. IV-STP, repeatability, St-Thi, and St-Att were expressed as coefficients of variation (%CV), while accuracy - defined here as phantom-based agreement with the reference BMD value - was reported as the complement of the mean absolute error (%MAE).
Results: %CV values for IV-STP and repeatability ranged from 0.50–1.49 and 0.61–1.24, respectively. St-Thi and St-Att tests showed %CV ranges of 0.93–6.65 and 0.32–3.87. Phantom-based accuracy varied between 81.48% and 92.75%. These results indicated heterogeneity and suboptimal stability and accuracy among some systems.
Conclusion: The proposed five-step QC protocol revealed performance variations among DEXA systems, including aspects not addressed in existing recommendations by international societies. This structured approach provides a more comprehensive framework for routine QC and can uncover latent deficiencies in system stability and accuracy, thereby enhancing diagnostic reliability and supporting regulatory standardization.
Keywords
Subjects

  1. Bono CM, Einhorn TA. Overview of osteoporosis: pathophysiology and determinants of bone strength. Eur Spine J. 2003 Oct;12 Suppl 2(Suppl 2):S90-6.
  2. El Maghraoui A, Koumba BA, Jroundi I, Achemlal L, Bezza A, Tazi MA. Epidemiology of hip fractures in 2002 in Rabat, Morocco. Osteoporos Int. 2005 Jun;16(6):597-602.
  3. Kanis JA, Johnell O, Oden A, Jonsson B, De Laet C, Dawson A. Risk of hip fracture according to the World Health Organization criteria for osteopenia and osteoporosis. Bone. 2000 Nov;27(5):585-90.
  4. Blake GM, Fogelman I. The role of DXA bone density scans in the diagnosis and treatment of osteoporosis. Postgrad Med J. 2007 Aug;83(982):509-17.
  5. Benmalek A, Sabatier JP. Comparison and cross-calibration of DXA systems: ODX-240 and Sophos L-XRA versus Hologic QDR-4500, for spinal bone mineral measurement. Translation of a reference database. Osteoporos Int. 1998;8(6):570-7.
  6. Genant HK, Guglielmi G, Jergas M. Bone densitometry and osteoporosis. Berlin, Heidelberg: Springer; 2012.
  7. Blake GM, McKeeney DB, Chhaya SC, Ryan PJ, Fogelman I. Dual energy x-ray absorptiometry: the effects of beam hardening on bone density measurements. Med Phys. 1992 Mar-Apr;19(2):459-65.              
  8. Mazess RB, Barden HS, Bisek JP, Hanson J. Dual-energy x-ray absorptiometry for total-body and regional bone-mineral and soft-tissue composition. Am J Clin Nutr. 1990 Jun;51(6):1106-12.
  9. Hansen MA, Hassager C, Overgaard K, Marslew U, Riis BJ, Christiansen C. Dual-energy x-ray absorptiometry: a precise method of measuring bone mineral density in the lumbar spine. J Nucl Med. 1990 Jul;31(7):1156-62.
  10. Davidsson L. Dual energy X ray absorptiometry for bone mineral density and body composition assessment. Vienna: International Atomic Energy Agency; 2010.
  11. Genant HK, Grampp S, Glüer CC, Faulkner KG, Jergas M, Engelke K, Hagiwara S, Van Kuijk C. Universal standardization for dual x-ray absorptiometry: patient and phantom cross-calibration results. J Bone Miner Res. 1994 Oct;9(10):1503-14.
  12. Khan AA, Colquhoun A, Hanley DA, Jankowski LG, Josse RG, Kendler DL, Lentle B, Leslie WD, Lewiecki EM, O'neill E, Robertson S, Syed ZA, Tanner SB, Webster D. Standards and guidelines for technologists performing central dual-energy X-ray absorptiometry. J Clin Densitom. 2007 Apr-Jun;10(2):189-95.
  13. Abbasi F: Comparison of bone mineral densitometry reproducibility among different dxa units and precision assessment by In-Vitro method. MSc thesis. Shiraz University, Department of Radiation Engineering; 2024.
  14. Steiger P. Standardization of measurements for assessing BMD by DXA. Calcif Tissue Int. 1995 Dec;57(6):469.
  15. Kim JS. Cross calibration of dual energy X-ray absorptiometry equipment for diagnosis of osteoporosis: between domestic manufacturers and global manufacturers. J Korean Soc Radiol. 2018;12(7):833–44.
  16. Nalda E, Mahadea KK, Dematteï C, Kotzki PO, Pouget JP, Boudousq V. Assessment of the Stratos, a new pencil-beam bone densitometer: dosimetry, precision, and cross calibration. J Clin Densitom. 2011 Oct-Dec;14(4):395-406.
  17. Laskey MA, Flaxman ME, Barber RW, Trafford S, Hayball MP, Lyttle KD, Crisp AJ, Compston JE. Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometers. Br J Radiol. 1991 Nov;64(767):1023-9.

Articles in Press, Accepted Manuscript
Available Online from 28 September 2026